163
Technical Data
10
Technical Data
Guaranteed data are specified for an ambient temperature of 23°C ± 5°
(73°F ± 9°), a power supply of 115/230V
AC
, and after a warm-up time longer
than 15 minutes.
Guaranteed data are valid for the period of one year after factory adjustment.
10.1
Mains Power Supply
10.2
Generator Output
10.3
Measurement Inputs
Mains power supply
Connection
Connector according to IEC 60320
Mains voltage
100 - 240V
AC
/ 50/60Hz / 6A
Instead of supplying the
CT Analyzer
from phase-
neutral (L1-N, A-N), it may also be supplied from
phase-phase (e.g. L1-L2, A-B). However, the
nominal voltage must not exceed 240V
AC
.
Mains fuses
2 x T6AH 250V,
(high-breaking capacity wire fuse 5 x 20mm)
Generator output
Output voltage / current
AC: 40V
rms
/ 5A
rms
max.
DC: 120V / 15A max.
Output power
400VA
rms
max.
Measurement input "Sec"
Voltage ranges
0 - 0.3 / 3 / 30 / 300V
AC
(auto ranging)
Accuracy
0.1% (guaranteed)
Summary of Contents for CT Analyzer
Page 1: ...CT Analyzer CT SB2 User Manual...
Page 3: ...CT Analyzer User Manual...
Page 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Page 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Page 50: ...CT Analyzer User Manual 48...
Page 60: ...CT Analyzer User Manual 58...
Page 80: ...CT Analyzer User Manual 78...
Page 162: ...CT Analyzer User Manual 160...
Page 172: ...CT Analyzer User Manual 170...
Page 188: ...CT Analyzer User Manual 186...
Page 194: ...CT Analyzer User Manual 192...
Page 195: ...CT SB2 User Manual...
Page 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Page 204: ...CT SB2 User Manual 10...
Page 206: ...CT SB2 User Manual 12...
Page 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Page 212: ...CT SB2 User Manual 18...
Page 226: ...CT SB2 User Manual 32...
Page 242: ...CT SB2 User Manual 48...
Page 248: ...CT SB2 User Manual 54...
Page 250: ...CT SB2 User Manual 56...
Page 252: ...CT SB2 User Manual 58...