CT Analyzer User Manual
182
905.xxx
Error [905.xxx] Over-temperature of power supply, wait until device has cooled down.
Reason:
Secondary side power supply has reached the warning temperature limit and
entered save mode to cool down.
Solution:
When this error is acknowledged as long it is active, the popup window is
closed and only the message in the status line remains active until the
overtemperature disappears. If the error is not acknowledged, the popup
window remains active.
Comment:
The status line displays the flashing message "905.xxx Overtemp.".
Safety action: All measurements are stopped and the error message is displayed in the status
line and a popup window. After acknowledgement, the software waits at least
1 minute before the test can be started again.
906.xxx
Error [906.xxx] Excess temperature of power supply, wait until device has cooled
down.
Reason:
The primary side power supply has reached its warning temperature limit and
entered save mode to cool down.
Solution:
When this error is acknowledged as long it is active, the popup window is
closed and only the message in the status line remains active until the
overtemperature disappears. If the error is not acknowledged, the popup
window remains active.
Comment:
The status line displays the flashing message "906.xxx Overtemp.".
Safety action: All measurements are stopped and the error message is displayed in the status
line and a popup window. After acknowledgement, the software waits at least
1 minute before the test can be started again.
908.xxx
Error [908.xxx] Device shut down.
Reason:
Internal power failure on the measurement interface module.
Solution:
The message is active as long as the internal power for the measurement
interface fails. Contact the next OMICRON service center if the error does not
disappear.
911.xxx
Error [911.xxx] Power supply error, contact next Omicron service center. Desired
voltage xxxx, measured voltage yyyy.
Reason:
One supply voltage on the measurement interface boards is out of tolerance.
Solution:
Contact your next OMICRON service center.
Comment:
It is no longer possible to work with the device.
912.xxx
Error [912.xxx] Excess internal temperature, wait until device has cooled down.
Reason:
Device-internal temperature at measurement interface > 75°C.
Solution:
Excessive heating of the device. Prevent the device from direct sunlight and
wait until it has cooled down. If the error occurs without previous exposure to
direct sunlight, the error is probably caused by a hardware error. Please
contact your next OMICRON service center.
Comment:
The status line displays the flashing message "912.xxx Overtemp.".
Safety action: All measurements are stopped and the error is displayed in the status line and
in a popup window. The error is active until the temperature falls below 60°C.
929.xxx
Error [929.xxx] Hardware failure, contact next Omicron service center.
Reason:
The circuit for reverse power dissipation does not work properly.
Solution:
Contact your next OMICRON service center.
Safety action: All measurements are stopped and the power supply is switched off.
Summary of Contents for CT Analyzer
Page 1: ...CT Analyzer CT SB2 User Manual...
Page 3: ...CT Analyzer User Manual...
Page 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Page 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Page 50: ...CT Analyzer User Manual 48...
Page 60: ...CT Analyzer User Manual 58...
Page 80: ...CT Analyzer User Manual 78...
Page 162: ...CT Analyzer User Manual 160...
Page 172: ...CT Analyzer User Manual 170...
Page 188: ...CT Analyzer User Manual 186...
Page 194: ...CT Analyzer User Manual 192...
Page 195: ...CT SB2 User Manual...
Page 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Page 204: ...CT SB2 User Manual 10...
Page 206: ...CT SB2 User Manual 12...
Page 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Page 212: ...CT SB2 User Manual 18...
Page 226: ...CT SB2 User Manual 32...
Page 242: ...CT SB2 User Manual 48...
Page 248: ...CT SB2 User Manual 54...
Page 250: ...CT SB2 User Manual 56...
Page 252: ...CT SB2 User Manual 58...