CT Analyzer User Manual
180
824.xxx
Error [824.xxx] An internal software error has occurred at address xxxxxxxH. To log
this error, make sure that a CF card is inserted and then press OK. Please send the file
\OMICRON\ErrorLog.xml on the CF card to the next OMICRON service center.
Reason:
Invalid result in a floating point operation (NaN).
Solution:
Contact your next OMICRON service center.
825.xxx
Warning [825.xxx] Size of file [FileName] >= 1Mbyte.
Reason:
The size of the file copied to the
CT Analyzer
Compact Flash card or virtual disc
is > 1 MB.
Solution:
Avoid transfer of large files > 1 MB to the
CT Analyzer
.
830.xxx
License < ... > is invalid.
Reason:
An invalid license entry was found in the license file
Omicron.lic
on the CF
Card.
Solution:
Specify a valid license key.
831.xxx
Setting value for < ... > is invalid. The default value of < ... > is used.
Reason:
Invalid item in file
Settings.inf
.
Solution:
The file
Settings.inf
in the
Omicron
directory on the CF card of the
CT Analyzer
has a formatting error in one of the parameters. Delete this file and
check the settings in the menu
Main Menu -> Settings -> Misc. Settings
. If
necessary, change the settings and store the file again.
832.xxx
Lifetime of relays near end. Contact the next OMICRON service center soon to arrange
a hardware service.
Reason:
The relays inside the
CT Analyzer
near the end of their lifetime (more than
1,000,000 measurements performed).
Solution:
Arrange a hardware service in order to replace the relays.
833.xxx
You are about to load a report with an invalid hash code. Some measurement values
may not be authentic.
Reason:
The file checksum is not valid.
Solution:
Check the authenticity of the report file loaded.
834.xxx
No valid license was found in this file!
Reason:
No valid license found.
Solution:
Check the license file. The license possibly does not correspond to the serial
number of the
CT Analyzer
.
835.xxx
Downgrade a firmware below version 3.53 is not supported.
Reason:
Attempt for a downgrade to a version number < 3.53.
Solution:
Make sure that die FW version number for the upgrade/downgrade is not below
3.53.
900.xxx
Error [900.xxx] Power supply error. Switch off the device and wait 1 minute to restart.
Reason:
The output power was larger than 350VA or the power supply is defective.
Solution:
Switch the
CT Analyzer
off, wait 1 minute and then switch it on again. If this
error occurs more frequently, you should contact your next OMICRON service
center.
Summary of Contents for CT Analyzer
Page 1: ...CT Analyzer CT SB2 User Manual...
Page 3: ...CT Analyzer User Manual...
Page 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Page 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Page 50: ...CT Analyzer User Manual 48...
Page 60: ...CT Analyzer User Manual 58...
Page 80: ...CT Analyzer User Manual 78...
Page 162: ...CT Analyzer User Manual 160...
Page 172: ...CT Analyzer User Manual 170...
Page 188: ...CT Analyzer User Manual 186...
Page 194: ...CT Analyzer User Manual 192...
Page 195: ...CT SB2 User Manual...
Page 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Page 204: ...CT SB2 User Manual 10...
Page 206: ...CT SB2 User Manual 12...
Page 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Page 212: ...CT SB2 User Manual 18...
Page 226: ...CT SB2 User Manual 32...
Page 242: ...CT SB2 User Manual 48...
Page 248: ...CT SB2 User Manual 54...
Page 250: ...CT SB2 User Manual 56...
Page 252: ...CT SB2 User Manual 58...