139
Using the Quick Test Feature
8.3
Performing Measurements with Quick Test
Measurements using
Quick Test
are not automatically stopped by the
CT Analyzer
. Such measurements have to be stopped manually by the
user.
Proceed as follows to perform measurements using the
Quick Test
feature:
1. Select
New Quick-Test
from the main menu and press the
OK
soft key to
open
Quick Test
. The display then shows the
CT-Quick
card.
Figure 8-2
CT-Quick
card
2. Select the measurement type and connect
your test object to the
CT Analyzer
correspondingly (see sections 8.4 to 8.10 below).
3. Define your measurement quantities.
4. Start and stop your measurement by pressing the
key.
5. After starting a measurement, the
Results
card is displayed automatically,
showing the measurement results.
8.4
Advanced Measurement
The Advanced measurement type offers complete multimeter measurement
functionality and free adjustment of all measurement quantities, such as AC or
DC signal output, voltage or current output, output frequency, measurement
mode (DC, RMS or frequency-selective) for the input, etc.
Using the Advanced measurement type it is also possible to use input "Sec" of
the
CT Analyzer
as a voltmeter for
measuring external voltages
up to
150V
RMS
with frequencies up to 4kHz.
Press the
Main
soft key in
any test card
Main Menu:
- New Quick-Test
OK
soft key
->
CT-Quick
card
How to get there:
I/0
Press the
Main
soft key in
any test card
Main Menu:
- New Quick-Test
OK
soft key
CT-Quick
card:
Measurement Type:
Advanced
How to get there:
Caution:
When using
Quick Test
, input "Sec" of the
CT Analyzer
is able to
measure voltages up to 150V
RMS
. Do not connect voltages above 150V or
even mains voltages to the
CT Analyzer
inputs! This will damage the
CT Analyzer
!
Summary of Contents for CT Analyzer
Page 1: ...CT Analyzer CT SB2 User Manual...
Page 3: ...CT Analyzer User Manual...
Page 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Page 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Page 50: ...CT Analyzer User Manual 48...
Page 60: ...CT Analyzer User Manual 58...
Page 80: ...CT Analyzer User Manual 78...
Page 162: ...CT Analyzer User Manual 160...
Page 172: ...CT Analyzer User Manual 170...
Page 188: ...CT Analyzer User Manual 186...
Page 194: ...CT Analyzer User Manual 192...
Page 195: ...CT SB2 User Manual...
Page 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Page 204: ...CT SB2 User Manual 10...
Page 206: ...CT SB2 User Manual 12...
Page 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Page 212: ...CT SB2 User Manual 18...
Page 226: ...CT SB2 User Manual 32...
Page 242: ...CT SB2 User Manual 48...
Page 248: ...CT SB2 User Manual 54...
Page 250: ...CT SB2 User Manual 56...
Page 252: ...CT SB2 User Manual 58...