141
Using the Quick Test Feature
f
Frequency of the output signal. Only available for AC voltage
or current output.
Possible values: Any value between 5.0 and 400.0Hz or soft
keys
16.7Hz
,
50Hz
,
60Hz
or
400Hz
.
Default: 50Hz
Measurement
Mode
Measurement mode.
Possible values:
DC
,
RMS
or
f-sel. filter
. Default: RMS
DC measurement:
Measures the DC part of the signal only.
RMS measurement:
Performs a standard true RMS measurement of the signal.
Frequency selective measurement:
Allows high precision measurement of the fundamental wave
of an applied signal. All frequencies outside the output
frequency plus/minus the filter bandwidth are supressed by at
least 110 dB. This measurement mode is deactivated for
output signal type DC.
See also "B-width" below.
B-width
Filter bandwidth for frequency selective measurement. Only
available for measurement mode
Freq. sel. filter
.
Possible values: 3Hz to 20Hz. Default: 6Hz
Depending on the filter bandwidth selected, the measurement
time can differ considerably. The typical time needed for a
frequency selective measurement depends on the filter
bandwidth as follows:
Filter bandwidth
Measurement time
3Hz approx.
5
seconds
6Hz
approx. 4 seconds
Load
Only available for AC signal output.
Setting the
Load
parameter to "on" and specifying the
winding resistance and the unsaturated inductance in the
fields
R
and
L
will prevent saturation of the connected CT.
The output generator of the
CT Analyzer
then keeps the DC
part of the output current at zero using a current regulator
permanently running during the output of AC signals.
Possible values:
On
or
Off
. Default: Off
Parameter
Description
Summary of Contents for CT Analyzer
Page 1: ...CT Analyzer CT SB2 User Manual...
Page 3: ...CT Analyzer User Manual...
Page 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Page 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Page 50: ...CT Analyzer User Manual 48...
Page 60: ...CT Analyzer User Manual 58...
Page 80: ...CT Analyzer User Manual 78...
Page 162: ...CT Analyzer User Manual 160...
Page 172: ...CT Analyzer User Manual 170...
Page 188: ...CT Analyzer User Manual 186...
Page 194: ...CT Analyzer User Manual 192...
Page 195: ...CT SB2 User Manual...
Page 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Page 204: ...CT SB2 User Manual 10...
Page 206: ...CT SB2 User Manual 12...
Page 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Page 212: ...CT SB2 User Manual 18...
Page 226: ...CT SB2 User Manual 32...
Page 242: ...CT SB2 User Manual 48...
Page 248: ...CT SB2 User Manual 54...
Page 250: ...CT SB2 User Manual 56...
Page 252: ...CT SB2 User Manual 58...