175
Error and Warning Messages
332.xxx
Note [332.xxx] I-sn has to be specified. Vb and VA are calculated dependent on I-sn.
Reason:
You tried to enter values for V
b
or VA without specifying the I
sn
value first.
Solution:
Enter a valid I
sn
value before specifying values for V
b
and VA.
333.xxx
Note [333.xxx] First define I-pn for this tap.
Reason:
You tried to enter a nominal burden (VA) value for a tap without specifying the
I
pn
for this tap first.
Solution:
Enter a valid I
pn
value for the tap before specifying the nominal burden (VA).
334.xxx
Note [334.xxx] Test cannot be started. Nominal burden (VA) should be >= 3.75VA.
Reason:
The nominal burden (VA) entered is lower than the limit specified in the device
settings (parameter "Min. VA at M cores Isn 5A").
Solution:
Enter a nominal burden (VA) higher than or equal to 3.75VA.
356.xxx
Warning [356.xxx] License: <All frequencies support> is missing! Only 60 Hz nominal
frequency is supported.
Reason:
The license required to use all frequencies is not available.
Solution:
Measurements are only possible for 60 Hz. Purchase a corresponding license
if you want to perform testing with other frequencies.
357.xxx
Note [357.xxx] License: <P cores support> is missing! Only metering CTs can be
measured.
Reason:
The license required to perform testing of protection cores is not available.
Solution:
Testing is only possible for metering cores. Purchase a corresponding license
if you want to perform testing of protection cores.
358.xxx
Warning [358.xxx] License: <M cores support> is missing! Only protection CTs can be
measured.
Reason:
The license required to perform testing of metering cores is not available.
Solution:
Testing is only possible for protection cores. Purchase a corresponding license
if you want to perform testing of metering cores.
360.xxx
Note [360.xxx] No class assessments will be made, because one or more of the settings
marked by a star ('?*' or '*') "are not defined.
Reason:
Some of the assessment-relevant settings are not defined.
Solution:
Make sure that all settings that are relevant for assessment (marked by '?*' or
'*') are defined.
400.xxx
Warning [400.xxx] You cannot make a multi-ratio measurement without a CT SB2 switch
box!
Reason:
No
CT SB2
is connected when starting a multi-ratio measurement.
Solution:
Connect the
CT SB2
to the
CT Analyzer
before starting a multi-ratio test.
401.xxx
Warning [401.xxx] CT SB2 switch box is disconnected!
Reason:
The
CT SB2
is not properly connected to the
CT Analyzer
.
Solution:
Restore proper connection between the
CT SB2
and the
CT Analyzer
and
restart your test.
402.xxx
Warning [402.xxx] CT SB2 switch box communication error.
Reason:
Communication between the
CT SB2
and the
CT Analyzer
failed.
Solution:
Restore proper connection between the
CT SB2
and the
CT Analyzer
and
restart your test.
Summary of Contents for CT Analyzer
Page 1: ...CT Analyzer CT SB2 User Manual...
Page 3: ...CT Analyzer User Manual...
Page 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Page 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Page 50: ...CT Analyzer User Manual 48...
Page 60: ...CT Analyzer User Manual 58...
Page 80: ...CT Analyzer User Manual 78...
Page 162: ...CT Analyzer User Manual 160...
Page 172: ...CT Analyzer User Manual 170...
Page 188: ...CT Analyzer User Manual 186...
Page 194: ...CT Analyzer User Manual 192...
Page 195: ...CT SB2 User Manual...
Page 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Page 204: ...CT SB2 User Manual 10...
Page 206: ...CT SB2 User Manual 12...
Page 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Page 212: ...CT SB2 User Manual 18...
Page 226: ...CT SB2 User Manual 32...
Page 242: ...CT SB2 User Manual 48...
Page 248: ...CT SB2 User Manual 54...
Page 250: ...CT SB2 User Manual 56...
Page 252: ...CT SB2 User Manual 58...