PV8344F4E013
45
(11)
Check the correctness of reference setting.
(12)
Check whether sampling interval is appropriate.
(13)
Check for injection of ambient light from fluorescent lamps or
the sun.
Failure in the
measurement of
wavelength direction
(1)
Refer to 5.3 Wavelength offset to conduct a offset.
Display of wavelength
in LUMINANCE
window is incorrect
(1)
Turn off the controller supply unit and reboot the computer.
Display of wavelength
in GRAPH window is
incorrect
(1)
Turn off the controller supply unit and reboot the computer.
Computer's
misbehaving
operation.
Failure in
measurement data.
(1)
Plug off and reinsert USB cable, then reboot the software.
(2)
Turn off the Control Power Suppy Box, then reboot the
computer.