PV8959-F4E005 25
LUMINANCE BY
PIXELS
Sets how to display a [LUMINANCE] window; the window
is displayed for each pixel of a photosensor when this
item is set to "Enable." It is displayed for each wavelength
when this item is set to "Disable." Usually, set this to
"Disable."
AUTO INCREMENT
Sets the storage data number of the spectral measured
value data during measurement; the empty data numbers
of the spectral measured values are automatically
searched every time measurement is executed when this
item is set to "Enable." The priority level of the search is
as follows: (1) the current data number, (2) a data number
higher than the current data number, and (3) the lowest
data number. Data is stored in the current data number
when there is no empty data number. When this item is
set to "Disable," data is stored in the specified data
number by the execution of measurement. Usually, set
this to "Disable."
FOCUS SIZE
Sets the index during the automatic adjustment of a
FOCUS window; usually, set this to 1000. The setting
range is an integer value of 100 to 5000.
FOCUS EXP
Sets the exposure time of a FOCUS camera; usually, set
this to 100. The setting range is an integer value of 1 to
525.
CT SIZE
Sets the index during the automatic adjustment of a FT
window; usually, set this to 1000. The setting range is an
integer value of 100 to 10000.
CT EXP
Sets the exposure time of a CT camera; usually, set this
to 250. The setting range is an integer value of 0 to 255.
FPS
Limits the loading speed of a camera; usually, set this to
20. The setting range is an integer value of 0 to 30.
Limitation is released when this item is set to 0.
SYSTEM COUNT1
This item is a parameter related to a FOCUS camera.
Usually, set this to 2. It is set according to the
performance of the computer.
SYSTEM COUNT2
This item is a parameter related to a FOCUS camera.
Usually, set this to 5. It is set according to the
performance of the computer.
USPM-RU3 CALIB
This item does not require change.