Novatest S.r.l.
Manual for Multichannel Ultrasonic Pile Integrity Tester
85
change accordingly after the channel and its corresponding pipe have
been selected.
Suggested (default) setup: Channel 1 corresponds to pipe 1,
Channel 2 corresponds to pipe 2, Channel 3 corresponds to pipe 3,
Channel 4 corresponds to pipe 4; in test, Channel 1 and 2 test Profile
1-2, Channel 1 and 3 test Profile 1-3, Channel 1 and 4 test Profile 1-4,
Channel 2 and 3 test Profile 2-3, Channel 2 and 4 test Profile 2-4,
Channel 3 and 4 test Profile 3-4.
NOTE: Different instruments have different number of
channels; the number of selected channels must be more than
1 but not more than number of pipes.
2
)
Profile status
Profile list of current pile displays just below the schematic
graph; the profile with “tested” is the profile which has been tested,
the optional profile is untested profile, the system will select profile
to be tested as default; select or cancel the check box in front of
profile name for manual change.
Displayed profile lists are different with different number of
pipes; please refer to Table 3.2 for details: