Manual for Multichannel Ultrasonic Pile Integrity Tester
135
2. Sampling length
Sample point Number of collected single-channel waveform
per time: there are 4 options (512, 1024, 2048 and 4096). The
default is 1024, and doo not input large value without special
requirement, otherwise it may negatively affect the speed in dynamic
collection)
3. Transmitting voltage
Select excitation voltage value when ultrasonic stimulated
emission transducer generates ultrasonic pulse, there are several
options including 65, 250, 500 and 1000. The default is 500V,
which is applicable to most test conditions. If other test conditions
remain unchanged, the higher the transmitting voltage is, the stronger
the receiving signal is.
It is suggested to select 65 or 250V when
testing the test block and small core sample.
4. Transmitting and receiving channels
Transmitting and receiving channel refers to the channel
connecting the transmitting and receiving transducer. The default
transmitting channel is channel 1 and the receiving channel is channel
2.
Channel 1, 2 and 3 can be selected as the transmitting channel of
U57OO. Only channel 1 can be selected as the transmitting channel
of U5600.