104
core sample should choose 65V or 250V.
4. Transmitting & Receiving Channels
Transmitting & Receiving Channels refer to channels connecting
with transmitting and receiving transducers, and the default
transmitting channel is channel 1, while the default receiving channel
is channel 2.
The transmitting channel of U5700 may choose among channel
1, channel 2 and channel 3, while the transmitting channel of U5600
only may choose channel 1.
5. Zero acoustic time
Zero acoustic time refers to the acoustic delay between
ultrasonic instrument and the transmitting and receiving transducers.
Zero acoustic time may be realized by manual input or press Zero
button to apply the zeroset function to automatically realize Zero
acoustic time. Please refer to section 3.6 for details).
NOTE: Parameters listed above only may be modified for
a new object; if the current object has been tested data, then
parameters listed above may not be modified except for Zero
acoustic time.
5.2.2.2 Calculation Parameter
Press Cal. Para. label, then switch to corresponding property
page (as shown in Figure 5.5), rule curve, aggregate type, coefficient