1-228
Confirmation and Check Sheet
7
7.1
Checking Procedure
The system is a FDA Class 3B and EN Class 3B laser product. The laser devices are FDA Class 3B and EN
Class 3b components. The final confirmation for the laser safety standards should be done after setting up.
Put a laser safety label on the prescribed place after confirming
.
(1) Laser intensity power
The recommended lasers can be put on the laser unit (see Table 5.1-1). Follow the process below and
confirm that calculated values satisfy the standard values to check laser intensities. If a calculated laser
intensity exceeds the standard value, contact the department in charge and do not ship the laser
product.
(1) Maximize each laser intensity.
(2) Remove all of movable ND filters from the optical path and adjust the argon laser to its maximum
power.
(2) Labels
Confirm the following labels put on the prescribed place (step 10 of “4.2.1 Assembling Each Part and
Attaching Them to the Microscope” in Chapter 4 of Part 1).
•
Laser radiation label.
•
Laser safety label.
(3) Check sheet
Fill out the necessary items on the check sheet and send it to the department in charge.