Chapter 4 Setting Up the Microscope
4.2 Setup Procedure
2-42
4.2.4
Attaching the Back Port Unit
Insert the back port unit into the guide on the rear of the microscope body and use the hexagonal
screwdriver (4-mm) to secure it from behind by tightening screws.
•
The following figure shows the back port unit being attached to the lower tier.
•
To attach the unit to the upper tier, attach a unit to be attached to the lower tier first and then attach the
one to the upper tier.
•
You must have installed the HUB controller when you use a HUB controller.
•
Attach the camera at the end.
Figure 4.2-20
It can be attached only to the lower tier when combined with the TIRF-PAU illuminator unit.
A separate back port extension kit is also required, which should be incorporated before attaching to the
microscope body.
Insert the back port unit body