![Nikon Eclipse Ti Series Setup Manual Download Page 291](http://html.mh-extra.com/html/nikon/eclipse-ti-series/eclipse-ti-series_setup-manual_1691710291.webp)
Chapter 4 Setting Up the Microscope
4.2 Setup Procedure
2-33
5
Assemble the remaining microscope parts.
Refer to the instruction manual included with the microscope main body before assembling the
remaining microscope parts.
6
Attach the laser safety cover (TIRF illuminator unit only).
LAS
ER
RA
DIA
TIO
N
AVO
ID E
XPO
SU
RE
TO
BE
AM
CLA
SS
3B L
ASE
R P
RO
DU
CT
AVO
ID E
XPO
SU
RE
LASER
LIGH
T IS
EMIT
TED
FROM
OBJ
EC
TIV
E A
PER
TURE
CAUTION – CLA
SS 3B
LASER RADIAT
ION WHEN OPE
N
AVOID EXPOSU
RE TO THE BEAM
VORSICHT – LA
SERSTRAHLUN
G
KLASSE 3B,
WENN ABDEC
KUNG GEÖFF
NET
NICHT DEM STR
AHL AUSSETZE
N
AVOID EXPOSU
RE
LASER LIGHT
IS EMITTED
FROM THIS A
PERTURE
CAUTION – CLA
SS 3B
LASER RADIAT
ION WHEN OP
EN
AVOID EXPOS
URE TO THE B
EAM
VORSICHT – LA
SERSTRAHLUN
G
KLASSE 3B,
WENN ABDEC
KUNG GEÖFF
NET
NICHT DEM ST
RAHL AUSSET
ZEN
CAUTION – CL
ASS 3B
LASER RADIA
TION WHEN O
PEN
AVOID EXPOS
URE TO THE B
EAM
VORSICHT – L
ASERSTRAHLU
NG
KLASSE 3B,
WENN ABDEC
KUNG GEÖFF
NET
NICHT DEM ST
RAHL AUSSET
ZEN
ON
OFF
6V30W
MAX.
12V100W
Figure 4.2-6
If you are using the TIRF illuminator unit, the motorized
shutter will not open unless the laser safety cover is
attached on the stage. Make sure it is attached in the
prescribed position.
1.
Place the laser safety cover body on the stage,
and secure it by tightening the 4 screws with the
Allen wrench.
2.
Connect the laser safety cover’s cable to the
safety-cover connector of the TI-LUSU shutter
unit.
3.
Attach the cover on top of the body.
The interlock will close the motorized shutter of
the laser unit if the cover is removed during use.
* In the case of the photo activation illuminator unit,
you must attach a specimen cover on the stage
when using episcopic illumination or laser light.
* Touching the interlocking part or getting it wet could
cause it to malfunction.
7
Attach the light shielding plate.
B
Figure 4.2-7
Using an Allen wrench, screw the light shielding plate
onto the eyepiece tube.
Laser safety cover
Upper cover
Body
Safety interlock (both sides)
Light shielding plate