Chapter 6 Connection between the Laser Unit and the Microscope
6.4 Installing the N-STORM Kit
1-227
Operating the N-STORM optical system
When the optical system for N-STORM microscopy is installed, the ND filter of the laser adapter, the STORM
slider, the lambda plate slider, and the cylindrical lens of the 3D-STORM port must be inserted or removed
according to the microscopy method as shown in the following table.
Table 6.4-1
Laser unit
Inverted microscope body
Microscopy
method
Laser adapter ND
filter
STORM slider
Lambda plate slider
3D-STORM port
cylindrical lens
Confocal
microscopy
OUT
Not used
Not used
Not used
Epifluorescent
microscopy
OUT
Not used
Not used
OUT
TIRF
microscopy
OUT OUT OUT OUT
2D
STORM
IN IN IN
OUT
3D
STORM
IN IN IN IN
IN
OUT
(The ND filter is placed within the optical path.)
(The ND filter is placed out of the optical path.)
Figure 6.4-2 IN state and OUT state of the laser adapter ND filter