4-44
BenchMike Pro 2025/2050 & Z-Mike Pro 4025/4050 Instruction Handbook: Setup
4.14.4 Screen Configuration
formation about these functions.
4.14.5 Scanner Configuration
From this page, you can send and receive information about
the configurat
ion of the BenchMike Pro’s scanner settings.
This information is typically only used by NDC Technologies
Service personnel.
DO NOT
use the Receive New function to transfer Scanner
Configuration settings from one BenchMike Pro to another
without contacting Customer Service first. Improper use of
this function may cause your BenchMike Pro to operate im-
properly. Note that you should not change the double differ-
entiation function. The gauge will not function properly if al-
tered.
4.14.6 Scan Errors Configuration
This page allows you to determine whether scanner errors are latched or
not. When enabled, all scan error conditions will persist until cleared by
the user.
Selecting Ignore Scan Errors disables scan error detection. This can re-
sult in erroneous measurements. Use this feature with caution.
4.14.7 International Configuration
From this page you can select UK English, US English,
French, German, or Spanish for all spelling, date, and num-
ber formats. Other international language support is availa-
ble. Contact NDC Technologies for details.
When Setup is complete, press the OK button to accept
changes.
At the Revision Screen, press
Exit
to return to the main
screen.