6-18
BenchMike Pro 2025/2050 & Z-Mike Pro 4025/4050 Instruction Handbook: Remote Setup
CONF.SER.PORT
Select a serial port number for setup and/or query. This
value can be set to 1. Examples:
CONF.SER.PORT:1
CONF.SER.PORT?
CONF.SER.SBITS
Request or specify the stop bits of the selected serial port.
This value can be either 1 or 2. Examples:
CONF.SER.SBITS:1
CONF.SER.SBITS?
IMPORTANT NOTE: Be careful to specify the correct baud rate when configuring the serial port. If you incorrectly
define the baud rate, you will be unable to communicate with the BenchMike Pro until this is corrected.
6.3.2
Getting Help
6.3.2.1
Help Commands: HELP
The following command lists all available help commands.
HELP
Request a listing of all supported PURL commands.
Examples:
HELP
6.3.3
Performing Measurements
6.3.3.1
Performing Measurements: MEAS
The MEASure command allows you to perform several operations related to measurements by BenchMike Pro.
MEAS.CLEARALL.NOW
This command will clear all data. Example:
MEAS.CLEARALL.NOW
MEAS.CLEARLATCHED.NOW
This command will clear all latched alarms. Example:
MEAS.CLEARLATCHED.NOW
MEAS.CLEARLAST.NOW
This command will clear the last measured data. Ex-
ample:
MEAS.CLEARLAST.NOW
MEAS.DATA
Enter a variable name followed by the equals sign to
set the value. Example:
MEAS.DATA:F1.NOMINAL=0.50
MEAS.DATA?F1.NOMINAL
MEAS.DEFINE.FORMAT
This command allows you to add a character string
to the .MEAS.LAST.MEAS command.
Example: