4-36
BenchMike Pro 2025/2050 & Z-Mike Pro 4025/4050 Instruction Handbook: Setup
The
Part Position
displays the vertical part position. Part Posi-
tion is the position of the part relative to the vertical center of
the scan. A positive value indicates that the part is above the
centerline, and a negative value indicates that the part is below
the centerline. Use this page to alter the title of this value to
something meaningful and relevant to your application, to dis-
play or not display the position on the main measurement dis-
play, and to transmit this value via the serial output port(s).
By default, the first line of data on the main measurement dis-
play contains the
Part Size
(diameter measurement). This la-
bel can be changed from SIZE to any other alphanumeric
character string by touching the SIZE box on this screen. This
measurement can be moved to another line or removed from
the display. The Size value can also be transmitted via the Se-
rial Output port.
Selecting the
SQC
command from the
Data
menu causes this
menu to be displayed.
NOTE: With the Serial Output drop-down box, you can trans-
mit these data items via the serial port. To set up the serial
port for data transmission, see Settings -Serial Output Format
and System-Serial for details.
These SQC values are computed:
Average
is the average Size reading, as computed
since the
last time data was cleared.
Count
is number of readings that have currently been
accumulated.
Maximum
is the largest Size reading detected since
data was last cleared.
Minimum
is the smallest Size reading detected since
data was last
cleared.
Diff/TIR
is difference between the Maximum value and
the Minimum
value.
Standard Deviation
is the standard deviation of the mean from the Nominal.
Batch
is the group of readings used to compute the above SQC values.