4-8
BenchMike Pro 2025/2050 & Z-Mike Pro 4025/4050 Instruction Handbook: Setup
The information received via the
Digital Inputs
port of the
BenchMike Pro is defined via the Digital Inputs page. The
trigger for each input can be set to: NO, NC, or Toggled.
Each of the inputs can be defined separately. These inputs
are:
Off
Go/Stop
Clear Data
Clear Last Data
Print Data
Send Data
Clear Latched Outputs
Floating Nominal
Advance Feature
Backup Feature
Emergency Stop
Zero Fixture 1
Zero Fixture 2
Home Fixture 1
Home Fixture 2
The
Digital Outputs
port pin assignment of the BenchMike
Pro is defined via the Digital Outputs page. There is one
standard output group plus two optional digital output
groups, of four functions each, from which to choose. Posi-
tive or negative polarity can be selected. Digital output
functions are user-assignable to specific digital outputs as
desired by the user.
NOTE: Digital outputs are open-collector outputs, with max-
imum specs of 35 V DC and 250mA.
These functions are:
Off
OK
Good
Reject High
Warning HIgh
Warning Low
Reject Low
Scan Error
Measuring
End of Part