BenchMike Pro 2025/2050 & Z-Mike Pro 4025/4050 Instruction Handbook: Remote Setup
6-13
CONF.SCANNER.FEAT.MEAS.LIMITS:OFF
CONF.SCANNER.FEAT.
MEAS.LIMITS.SEL
Request or specify the type of measurement limit to be
set by the .TOLVAL command. Applies only to this fea-
ture. Options are: +REJ, +WARN, -WARN, or -REJ. Exam-
ples:
CONF.SCANNER.FEAT.MEAS.LIMITS.SEL:+REJ
CONF.SCANNER.FEAT.MEAS.LIMITS.SEL?
CONF.SCANNER.FEAT.
MEAS.LIMITS.TOLVAL
Request or specify the measurement tolerance value for
the currently selected limit type. This is the amount of
deviation from the nominal. The limit values used in
other commands in this table are automatically comput-
ed with this value. Examples:
CONF.SCANNER.FEAT.MEAS.LIMITS.TOLVAL:0.500
CONF.SCANNER.FEAT.MEAS.LIMITS.TOLVAL?
CONF.SCANNER.FEAT.
MEAS.NOM
Request or specify the nominal part size for limits
checking. Examples:
CONF.SCANNER.FEAT.MEAS.NOM:0.5000
CONF.SCANNER.FEAT.MEAS.NOM?
CONF.SCANNER.FEAT.
MEAS.PART
Turn “part mode” ON or OFF for this feature. When ON,
the BenchMike Pro will stop measuring the part once it
is removed from the measurement area. Examples:
CONF.SCANNER.FEAT.MEAS.PART:ON
CONF.SCANNER.FEAT.MEAS.PART:OFF
CONF.SCANNER.FEAT.
MEAS.AVG.VAL
Request or specify the time period for measurements to
be averaged.The time unit used is determined by
.CONF.SCANNER.FEAT.MEAS.AVG. Examples:
CONF.SCANNER.FEAT.MEAS.AVG.VAL:200
CONF.SCANNER.FEAT.MEAS.AVG.VAL?
CONF.SCANNER.FEAT.
MEAS.METHOD
Request or specify the type of data sampling: SINGLE,
CONTINUOUS, or PART. Examples:
CONF.SCANNER.FEAT.MEAS.METHOD:CONTINUOUS
CONF.SCANNER.FEAT.MEAS.METHOD?
CONF.SCANNER.FEAT.
MEAS.TYPE
Request or specify the type of measurements to be tak-
en for this feature. These options are listed under
. Examples:
CONF.SCANNER.FEAT.MEAS.TYPE:DIAMETER
CONF.SCANNER.FEAT.MEAS.TYPE?
CONF.SCANNER.FEAT.SEL
Select the number of the active feature. The number can
range from 1 to 32. Examples:
CONF.SCANNER.FEAT.SEL:1
CONF.SCANNER.FEAT.SEL?