BenchMike Pro 2025/2050 & Z-Mike Pro 4025/4050 Instruction Handbook: Setup
4-37
The
SQC Setup
menus for each of the SQC parameters are identical except for the title. Use these pages to
change the label of the parameter or change its location on the main measurement page. You can also decide to
transmit these values through the serial output. These parameters are: Average, Count, Diff/TIR, Maximum, Min-
imum, and Standard Deviation.
The
Optical Plate Temperature
displays the optical plate
temperature as entered by the user. When Thermal Compen-
sation is enabled, this value affects the measurement.
The
Motherboard Temperature
displays the motherboard
temperature as entered by the user. When Thermal Compen-
sation is enabled, this value affects the measurement.
The
Actual Temperature
displays the part temperature as
entered by the user. When Thermal Compensation is enabled,
this value affects the measurement.