3-23
Method 2:
Tools required:
BIS simulator (with flexible cable)
BIS cable (PN: 6800-30-50761
)
1. Connect the BIS sensor with the BIS simulator and select [
BIS Setup
]
→
[
BIS Sensor
Check
] to perform a cyclic impedance check.
2. After the cyclic impedance check is finished, check that the result for each electrode is
pass.
3. Check the EEG wave and BIS numeric displayed on the screen.
3.4.13 RM Test
Tool required:
Gas source
Ventilator (calibrated)
Adult artificial lung
(
PN: 040-000744-00
)
RM connector (PN: 6800-20-50328
)
Adult/pediatric flow sensor (PN: 0010-30-42678)
Follow this procedure to perform the test:
1. Connect the equipment as shown above. Make sure that the blue sensing tube on the
flow sensor is connected with the artificial lung.
2. Set
[
Patient Cat.
] to [
Adu
]. In the [
RM Setup
] menu, select [
Sensor Type
] according
to the used sensor and set [
Ventilation Mode
] to [
Mechanical
].
3. Enter
the
[
RM Setup
] menu and select [
Calibrate
>>]. Input the constant marked on
the sensor and calibrate the flow sensor.
4. Configure the ventilator as follows: TV=500 ml, RR =20 rpm, I:E=1:2.
Ventilator
Artificial lung
Monitor
Flow sensor
Summary of Contents for BeneView T5
Page 1: ...DPM 6 Patient Monitor Service Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 12: ...6 FOR YOUR NOTES ...
Page 16: ...1 4 FOR YOUR NOTES ...
Page 32: ...2 16 FOR YOUR NOTES ...
Page 68: ...3 36 FOR YOUR NOTES ...
Page 80: ...4 12 FOR YOUR NOTES ...
Page 126: ...5 46 FOR YOUR NOTES ...
Page 133: ...6 7 6 3 3 12 1 Screen Assembly with anti glare screen Exploded View ...
Page 163: ...6 37 6 6 7 BIS Module Exploded View ...
Page 180: ...6 54 FOR YOUR NOTES ...
Page 204: ...A 14 FOR YOUR NOTES ...
Page 205: ......