5-21
2. Unscrew the 4 M3
×
6 screws and take out the main board assembly.
NOTE
z
Since the main board assembly is connected with the power module via a butt
socket and the CPU radiator may be adhered to it, the main board assembly should
be removed with force. Take care not to damage the butt socket.
3. Unscrew the 4 M2.5
×
6 screws on the CPU board and vertically separate the CPU board
and the mother board. Take care not to damage the butt socket.
Ring