4.6.11 Software Upgrade Problems
Symptoms
Possible Cause
Corrective Action
Bootstrap upgrade
fails
Power failure or
unintended power off
during bootstrap upgrade
Return the CPU board to factory for repair.
Incorrect network
connection
1. Check that network connector, not CIS
connector, on the patient monitor is used.
2. Make sure that the hub or switch run
normally. Check that net twines are of the
right type and have been connected correctly.
Wrong upgrade package
has been downloaded
Upgrade package shall be .pkg files. Select
package according to system requirement.
Program upgrade fails
Incorrect IP address
configuration
Configure a fixed IP address in range C as
specified for the patient monitor. We
recommend not to upgrade a program when
the patient monitor is connected to a network
with multiple PCs.
4.6.12 Technical Alarm Messages
Please refer to the Operator’s manual.
4.6.13 M51A Self Test Information
New MPM module applies the integrative parameter board (ECG ASIC).
MPM Selftest Item
Test Value
Test Value (New
MPM module)
Corrective Action
DSP selftest
information
Not F
7024 selftest
information
Not 7F
2131 selftest
information
Not 1F
ECG module selftest
information
Normal value
:
7 for 3/5
lead module; FF for
12-lead module
Not FF
Replace the module
Summary of Contents for BeneView T5
Page 1: ...DPM 6 Patient Monitor Service Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 12: ...6 FOR YOUR NOTES ...
Page 16: ...1 4 FOR YOUR NOTES ...
Page 32: ...2 16 FOR YOUR NOTES ...
Page 68: ...3 36 FOR YOUR NOTES ...
Page 80: ...4 12 FOR YOUR NOTES ...
Page 126: ...5 46 FOR YOUR NOTES ...
Page 133: ...6 7 6 3 3 12 1 Screen Assembly with anti glare screen Exploded View ...
Page 163: ...6 37 6 6 7 BIS Module Exploded View ...
Page 180: ...6 54 FOR YOUR NOTES ...
Page 204: ...A 14 FOR YOUR NOTES ...
Page 205: ......