3-35
respectively.
Nurse call relay performance test
The relay contacts are close when an alarm occurs.
Yes No
Analog output performance test
The waves displayed on the oscillograph are identical with those
displayed on the monitor.
Yes No
Electrical safety tests
Refer to
A Electrical Safety Inspection
. All the electrical safety
tests should be passed.
Yes No
Touchscreen calibration
The touchscreen is calibrated successfully.
Yes No
Recorder check
The recorder can print ECG waves correctly and the printout is
clear.
Yes No
Set the recorder to some problems such as out of paper, paper
jam, etc. the monitor gives corresponding prompt messages.
After the problem is removed, the recorder is able to work
correctly.
Yes No
Automatic alarm recording for each parameter functions
correctly when parameter alarms occur.
Yes No
Network print test
The network printer can print out ECG reports correctly.
Yes No
Battery check
The monitor can operates correctly from battery power when an
AC power failure accidentally occurs.
Yes No
T5 patient monitor can operate independently on a single battery.
Yes No
The operating time of the battery meets the product
specification.
Yes No
Test conclusion:
Tested by: Test Date:
Summary of Contents for BeneView T5
Page 1: ...DPM 6 Patient Monitor Service Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 12: ...6 FOR YOUR NOTES ...
Page 16: ...1 4 FOR YOUR NOTES ...
Page 32: ...2 16 FOR YOUR NOTES ...
Page 68: ...3 36 FOR YOUR NOTES ...
Page 80: ...4 12 FOR YOUR NOTES ...
Page 126: ...5 46 FOR YOUR NOTES ...
Page 133: ...6 7 6 3 3 12 1 Screen Assembly with anti glare screen Exploded View ...
Page 163: ...6 37 6 6 7 BIS Module Exploded View ...
Page 180: ...6 54 FOR YOUR NOTES ...
Page 204: ...A 14 FOR YOUR NOTES ...
Page 205: ......