3-2
3.1.2 Test Report
Upon completion of the tests, the table of preventative maintenance test reports and the table
of maintenance test reports in this chapter should be kept properly.
3.1.3 Preventative Maintenance
Below are preventative maintenance tests which need to be performed on the monitor. See
the following sections for detailed maintenance procedures.
Visual inspection
NIBP test and calibration
Microsteam and Sidestram CO
2
test and calibration
AG test and calibration
3.1.4 Recommended Frequency
Check/Maintenance Item
Frequency
Preventative Maintenance Tests
Visual inspection
1. When first installed or reinstalled.
Pressure check
Leakage test
NIBP test
Calibration
Leakage test
Performance test
Sidestream and
Microstream
CO
2
tests
Calibration
Leakage test
Performance test
AG tests
Calibration
1. If the user suspects that the measurement is
incorrect.
2. Following any repairs or replacement of relevant
module.
3.At least once a year is recommended.
Performance Tests
Performance test
ECG test and
calibration
Calibration
Resp
performance test
/
1. If the user suspects that the measurement is
incorrect.
2. Following any repairs or replacement of relevant
module.
3. At least once every two years.
Summary of Contents for BeneView T5
Page 1: ...DPM 6 Patient Monitor Service Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 12: ...6 FOR YOUR NOTES ...
Page 16: ...1 4 FOR YOUR NOTES ...
Page 32: ...2 16 FOR YOUR NOTES ...
Page 68: ...3 36 FOR YOUR NOTES ...
Page 80: ...4 12 FOR YOUR NOTES ...
Page 126: ...5 46 FOR YOUR NOTES ...
Page 133: ...6 7 6 3 3 12 1 Screen Assembly with anti glare screen Exploded View ...
Page 163: ...6 37 6 6 7 BIS Module Exploded View ...
Page 180: ...6 54 FOR YOUR NOTES ...
Page 204: ...A 14 FOR YOUR NOTES ...
Page 205: ......