3.2.5 Preventative maintenance test report
Customer name
Customer address
Servicing person
Servicing company
Equipment under test
(EUT)
Model of EUT
SN of EUT
Hardware version
Software version
Test equipment
Model/No.
Effective date of calibration
Test items
Test records
Test
results(Yes/No)
Visual inspection
The case, display screen, buttons, knob, SMR, modules, power
cord, wall-mount bracket and accessories have no obvious signs
of damage.
Yes No
The external connecting cables are not frayed and the connector
pins are not loose and bent.
Yes No
The external connectors are not loose or their pins are not bent.
Yes No
The safety labels and data plate are clearly legible.
Yes No
NIBP test
The difference is within ±3 mm when 0, 50 or 200 mmHg is set
for NIBP accuracy test.
Yes No
There is no leakage with NIBP, or the manual leakage test result
does not exceed 6mmHg/min.
Yes No
Sidestream CO
2
test