5-36
5.5.2 Disassembling CO
2
Module
1. Remove the four contact screws on the back and the two snap plates at the bottom and
then unscrew the two M3×6 screws by referring to the procedure as described in
5.5.1
Disassembling the ICG Module
.
2. Remove the module’s front panel by referring to the procedure as described in
5.5.1
Disassembling the ICG Module
. Then, press the two clips on the inner housing about
1mm inwards to separate the inner and outer housing. Then, take off the outer housing.