3-20
3.4.8 Mainstream CO
2
Tests
NOTE
z
Select [Main Menu]
→
[Maintenance >>]
→
[User Maintenance >>]
→
enter the
required password
→
[Maintain CO
2
], make sure that the setting of [Barometric
Pressure] is correct before performing mainstream CO
2
tests.
Tools required:
A steel gas cylinder with 6±0.05% CO
2
A steel gas cylinder with compressed air or N
2
(with standard concentration)
Two 3-way valves (power supply controlled)
Flowmeter
Power supply
Tube
Follow this procedure to perform the test:
1. Wait
until
CO
2
warmup is finished and then select [
Start Zero Cal.
]
from [
CO
2
Setup
]
menu to start a zero calibration. If the zero calibration fails, the prompt message [
CO
2
Zero Failed
] is displayed. Otherwise, the baseline of waveform recovers to zero.
2 Set
[
Apnea Delay]
to 10 s in the [
Adjust CO
2
Limits
] menu.
3 Blow
to
the
CO
2
sensor to generate a CO
2
waveform and then place the sensor in the air.
Check if the alarm message [
CO
2
Apnea]
is displayed on the screen.
4
Connect the test system as follows
Summary of Contents for BeneView T5
Page 1: ...DPM 6 Patient Monitor Service Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 12: ...6 FOR YOUR NOTES ...
Page 16: ...1 4 FOR YOUR NOTES ...
Page 32: ...2 16 FOR YOUR NOTES ...
Page 68: ...3 36 FOR YOUR NOTES ...
Page 80: ...4 12 FOR YOUR NOTES ...
Page 126: ...5 46 FOR YOUR NOTES ...
Page 133: ...6 7 6 3 3 12 1 Screen Assembly with anti glare screen Exploded View ...
Page 163: ...6 37 6 6 7 BIS Module Exploded View ...
Page 180: ...6 54 FOR YOUR NOTES ...
Page 204: ...A 14 FOR YOUR NOTES ...
Page 205: ......