3-21
In the figure above,
1
A steel gas cylinder with 6±0.05% CO
2
2 Flowmeter
3
3-way valve (power supply controlled)
4
Open to air
5
Power supply (controlling two 3-way valves)
6
Compressed air or N
2
with standard concentration
7 Mainstream
CO
2
sensor
8 Patient
monitor
9
Tube (preventing back flow)
5
Adjust the power supply and turn on/off 3-way valves to ensure that that only one
cylinder is connected to the Mainstream CO
2
sensor via the 3-way valves at one time
and the flowmeter reading is stable and within 2 and 5L/min.
6
Switch between the two cylinders to connect Mainstream CO
2
sensor at an intervals of 6
to 10s and check if the displayed CO
2
value is within 6.0
±
0.3
%
.
3.4.9 Sidestream and Microstream CO
2
Module Tests
See section
3.2.3 Sidestream and Microstream CO
2
Module Tests.
3.4.10 AG Tests
See section
3.2.4 AG Tests.