5.5 Disassembling Modules
WARNING
z
For a reassembled module, a patient leakage current test must be performed
before it is used again for patient monitoring.
z
Make sure the monitor is off before disassembling the modules.
The monitor supports three types of modules, single-slot modules, 2-slot modules and 3-slot
modules. The following part describes two typical disassembling procedures.
5.5.1 Disassembling the ICG Module
1. Unscrew the two contact screws and
ф
3 spring washers on the back of the module with
a dedicated contact spanner.
Contact screw