3-25
3.5 Nurse Call Relay Performance Test
Tools required:
Oscilloscope
1. Connect the nurse call cable to the Nurse Call Connector of the patient monitor.
2. Enter
[
Demo
] mode. Then, select [
Main Menu
]
→
[
Maintenance >>
]
→
[
User
Maintenance >>
]
→
enter the required password
→
[
Others >>
]
.
3. In
the
[
Others >>
] menu, select [
Nurse Call Setup >>
] and then in the [
Nurse Call
Setup
] menu, select all optional [
Alm Lev
] and [
Alm Cat.
], and set
[Contact Type]
to
[Normally Open]
4. In
[
Nurse Call Setup >>
] menu, set [
Signal Type
] to [
Pulse
]. Make the monitor to
generate an alarm and check that the oscillograph displays positive pulses of 1s width
when there is an alarm.
5. In
[
Nurse Call Setup >>
] setup menu, set [
Signal Type
] to [
Continuous
].Make the
monitor to generate an alarm and check that the oscillograph outputs continuous high
level when there is an alarm.
3.6 Analog Output Performance Test
Tool required:
Medsim300B patient simulator, or MPS450, or equivalent equipments
Oscillograph
Connect the patient simulator to the monitor using an ECG or IBP cable and connect the
oscillograph to the Auxiliary Output Connector of the patient monitor. Verify that the waves
displayed on the oscillograph are identical with those displayed on the monitor.
Summary of Contents for BeneView T5
Page 1: ...DPM 6 Patient Monitor Service Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 12: ...6 FOR YOUR NOTES ...
Page 16: ...1 4 FOR YOUR NOTES ...
Page 32: ...2 16 FOR YOUR NOTES ...
Page 68: ...3 36 FOR YOUR NOTES ...
Page 80: ...4 12 FOR YOUR NOTES ...
Page 126: ...5 46 FOR YOUR NOTES ...
Page 133: ...6 7 6 3 3 12 1 Screen Assembly with anti glare screen Exploded View ...
Page 163: ...6 37 6 6 7 BIS Module Exploded View ...
Page 180: ...6 54 FOR YOUR NOTES ...
Page 204: ...A 14 FOR YOUR NOTES ...
Page 205: ......