Customizing the Analyzer Software
5-17
Figure 5-29 Setting unit and count time
Selecting
the
“
Count
” group
Press [F1] to select the “
Count
” group, as Figure 5-29 shows.
Selecting
parameter
units
This analyzer provides multiple units for certain parameters. Refer to Table 5-2 for all the
selectable units of all parameters. The 19 parameters are divided into 11 groups based on
their units and you can only select unit for the first parameter of a group. Pay special attention
to the HGB group, which includes HGB, MCHC and MCH. When you select g/L or g/dL as the
unit of HGB, the default unit for MCH is pg; when you select mmol/L as the unit of HGB, the
default unit of MCH is fmol.
Table 5-2 Units of Parameters
Parameter Display
format Unit
Remarks
***.*
10
9
/L Default
***.* 10
3
/uL /
****
10
2
/uL /
WBC
Lymph#
Mid#
Gran#
***.* /nL /
**.*
Lymph%
Mid%
Gran%
.***
%
Default
***
g/L
Default
**.* g/dL /
HGB, MCHC
**.* mmol/L
/