Maintaining Your Analyzer
10-13
Figure 10-15 A message box to confirm the cleaning
10.3.8 Lyse Test
z
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
z
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
In case of any abnormal WBC counts or histograms, you can perform the “
Lyse test
”
procedure to check whether the lyse can be dispensed properly.
Follow the steps given below to do so:
1. Unscrew and remove the retaining screws with hands or screwdrivers (pointed by the
arrows shown in Figure 10-16) on the right plate;
Figure 10-16 Removing the two screws
2. Follow the arrow shown in Figure 10-17 to push and remove the right plate;