Customizing the Analyzer Software
5-19
Figure 5-30 Selecting parameter unit
Setting
count
time
If the WBC or RBC count time is inappropriately set, the system may give false alarms for
clogs or bubbles. When this happens, follow the steps given below to change the WBC or
RBC count time. Refer to the actual count time (see
Chapter 10.5.1
for details) when editing
the count time. Follow the steps given below to set the count time.
1. Enter the administrator password as introduced in
Chapter 5.2.1
;
2. At the “
Settings
” screen, press [F1] to select the “
Count
” group;
3.
SELECT
“
WBC
” or “
RBC
”, as Figure 5-31 shows;
4.
ENTER
the desired digits.
Figure 5-31 Setting count time
Exiting the “
Count
” group