Appendices
B-3
B.5.2 Normal background
Parameter Background
result
WBC
≤
0.3
×
10
9
/ L
RBC
≤
0.03
×
10
12
/ L
HGB
≤
1 g / L
HCT
≤
0.5 %
PLT
≤
7
×
10
9
/ L
B.5.3 Linearity range
Parameter Linearity
range
WBC (10
9
/L)
0.0 to 99.9
RBC (10
12
/L)
0.0 to 9.99
HGB (g/L)
0 to 300
PLT (10
9
/L)
0 to 999
B.5.4 Carryover
Parameter Carryover
WBC
≤
0.5 %
RBC
≤
0.5 %
HGB
≤
0.5 %
PLT
≤
1 %
B.5.5 Reproducibility (with normal-level control)
Parameter Condition
CV(%)
WBC (
×
10
9
/L)
4.0 to 15.0
≤
3.0
RBC (
×
10
12
/L)
3.00 to 6.50
≤
2.0
HGB (g/L)
100 to 180
≤
2.0
MCV (fL)
70 to 100
≤
1.0
PLT(
×
10
9
/L)
200 to 500
≤
5.0
B.6 Input/Output
Device
z
Use the specified devices only.