Maintaining Your Analyzer
10-5
the diluent in the tubing is contaminated; or
you have installed a new container of diluent without shutting the analyzer.
At the “
Maintenance
” screen,
SELECT
“
Diluent prime
” to prime the tubing and the priming
progress will be displayed at the bottom of the screen, Figure 10-3 shows. When the priming
is done, the screen will return to the initial state.
Figure 10-3 Priming diluent screen
10.3.2 Rinse Prime
z
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
z
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
z
After installing new reagents, let the reagents stand for a while before using
them.
z
After installing a new container of diluent, rinse or lyse, do a background
check to ensure the background results are normal.
Summary of Contents for BC-2800
Page 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Page 2: ......
Page 12: ......
Page 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Page 62: ......
Page 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Page 92: ......
Page 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Page 196: ......
Page 248: ......
Page 252: ......
Page 266: ......
Page 284: ...Appendices E 10 Hemoglobin Concentration ...
Page 286: ...P N 2800 20 28795 2 0 ...