Operating Your Analyzer
6-8
z
Keep dust from the prepared diluent.
z
When
preparing
a
prediluted
sample
,
only
use lint-free
tissue
to
wipe
the
external
wall
of
the
capillary
tube
;
do
not
use
cotton
balls
.
z
After mixing the capillary sample with the diluent, wait 5 minutes before
running the sample.
z
Run the prediluted samples within 30 minutes after the mixing.
z
Mix any sample that has been prepared for a while before running it.
z
Evaluate predilute stability based on your laboratory’s sample population
and sample collection techniques or methods.