Troubleshooting Your Analyzer
11-8
Recorder too hot Recorder head too hot.
1. Stop using recorder and cool it for 10
minutes and then print again;
2. If the problem remains, shut down the
analyzer and restart it again, and then
print again;
1. If the problem remains, contact
Mindray customer service department.
Lyse out
Insufficient lyse or wrong lyse
volume setting.
1. Check if there is sufficient lyse left. If
so, access “
Setup
→
Settings
→
Reagents
” and adjust the remaining
lyse volume as instructed by
Chapter
5.3.1
;
2. If not, change a new container of lyse
as instructed by
Chapter 4.4.1
.
Diluent expired
Expired diluent or wrong
expiration setting
1. Check if the diluent has expired. If so,
change a new container of diluent as
instructed by
Chapter 4.4.1
;
2. If not, reset the expiration date as
instructed in
Chapter 5.3.1
.
Rinse expired
Expired rinse or wrong
expiration setting
1. Check if the rinse has expired. If so,
change a new container of rinse as
instructed by
Chapter 4.4.1
;
2. If not, reset the expiration date as
instructed in
Chapter 5.3.1
.
Lyse expired
Expired lyse or wrong
expiration setting
1. Check if the lyse has expired. If so,
change a new container of lyse as
instructed by
Chapter 4.4.1
;
2. If not, reset the expiration date as
instructed in
Chapter 5.3.1
.
Vacuum filter
error
The air inside the vacuum
chamber is not extracted
within the given time.
1. Enter
“
Service
→
Self–test
→
Tubing
”
to test the filter as instructed
in
Chapter 10.5.1
;
2. If the test result is normal, the error will
be removed;
3. If the problem remains, replace the
filter as instructed by
Chapter 10.11
;
4. If the problem still remains after a new
filter has been installed, contact
Mindray customer service department
Summary of Contents for BC-2800
Page 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Page 2: ......
Page 12: ......
Page 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Page 62: ......
Page 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Page 92: ......
Page 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Page 196: ......
Page 248: ......
Page 252: ......
Page 266: ......
Page 284: ...Appendices E 10 Hemoglobin Concentration ...
Page 286: ...P N 2800 20 28795 2 0 ...