Maintaining Your Analyzer
10-9
10.3.6 Probe Cleanser Cleaning
z
Samples, controls, calibrator and waste are potentially infectious. Wear
proper personal protective equipment (e.g. gloves, lab coat, etc.) and follow
safe laboratory procedures when handling them in the laboratory.
z
The probe cleanser is corrosive. Wear proper personal protective equipment
(e.g. gloves, lab coat, etc.) and follow safe laboratory procedures when
handling them in the laboratory.
You can soak the bath and fluidic lines with the probe cleanser, an alkaline detergent, by
performing the “
Probe cleanser cleaning
” procedure. If your analyzer is to run 24 hours a
day, you should perform this procedure.
Follow the steps given below to do so:
1. At the “
Maintenance
” screen, move the cursor to
“
Probe cleanser cleaning
”;
2. Present the cleanser to the probe and press [ENTER] to aspirate the cleanser. When you
hear the beep and the sample probe is out of the bottle, remove the cleanser. The
analyzer will start priming process, as Figure 10-8 shows;
Figure 10-8 Priming bath and fluidic lines
3. When the priming is done, the analyzer will start the 5-minute soaking process, as Figure
10-9 shows and you may press [ENTER] to stop it before the time is due. Note that a
Summary of Contents for BC-2800
Page 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Page 2: ......
Page 12: ......
Page 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Page 62: ......
Page 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Page 92: ......
Page 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Page 196: ......
Page 248: ......
Page 252: ......
Page 266: ......
Page 284: ...Appendices E 10 Hemoglobin Concentration ...
Page 286: ...P N 2800 20 28795 2 0 ...