Using the QC Programs
8-4
shows.
Figure 8-5 Entering “Exp. Date”
Entering the expected results (mean) and limits (range)
ENTER
the expected results (mean) and limits (range) respectively into the “
Mean
”
and
“
Range
” boxes of the parameters to be included in the QC analysis, as Figure 8-6 shows.
Figure 8-6 Entering “Mean” of RBC
z
Refer to the instructions of use of the control for information on the lot
number, expiration date, open-vial stability days, expected results and limits.
z
The entered expiration date should be either the expiration date printed on
the labeling or the open-vial expiration date. It is earlier.