Operating Your Analyzer
6-7
refrigerated samples at room temperature for at least 30 minutes before
running them.
z
Mix any sample that has been prepared for a while before running it.
6.6.2 Prediluted Samples
Collect and handle the prediluted sample as follows:
1. Press [MENU] and
SELECT
”
Count
” to enter the ”
Count
” screen;
2. Press [MODE] to select prediluted analysis mode (any mode preceded by
Prediluted
);
3. Press [DILUENT] and a message box will pop up to instruct you how to dispense the
diluent into the sample tube, as Figure 6-1 shows;
Figure 6-1 A message box showing you how to dispense diluent
4. Present a clean sample tube to the sample probe and make sure the tube is tilted
towards the probe, as Figure 6-2 shows, to avoid spills and bubbles. Press the aspirate
key to dispense 0.7mL of diluent (the dispensing volume is controlled by the analyzer)
into the tube;
Figure 6-2 How to dispense diluent
5. When the dispensing is finished, press [ENTER] to close the message box;
6. Add 20µL of capillary blood to the diluent and shake the tube to mix the sample.
Summary of Contents for BC-2800
Page 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Page 2: ......
Page 12: ......
Page 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Page 62: ......
Page 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Page 92: ......
Page 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Page 196: ......
Page 248: ......
Page 252: ......
Page 266: ......
Page 284: ...Appendices E 10 Hemoglobin Concentration ...
Page 286: ...P N 2800 20 28795 2 0 ...