Maintaining Your Analyzer
10-19
10.3.12 Clean Probe Wipe
After being used for a long time, the bottom of the probe wipe may be contaminated by blood
and the inside of the probe wipe may also be contaminated by the dirt sucked in. So you need
to clean the probe wipe regularly.
Follow the steps given below to do so:
1. Present the probe cleanser to the sample probe;
2.
SELECT
“
Clean Probe Wipe
” to start the aspiration;
3. When you hear the beep and the sample probe is out of the bottle, remove the cleanser;
4. Unscrew and remove the retaining screws with hands or screwdrivers (pointed by the
arrows shown in Figure 10-25) on the right plate;
Figure 10-25 Removing the two screws
5. Follow the arrow shown in Figure 10-26 to push and remove the right plate;
Figure 10-26 Removing right plate
6. Follow the instructions displayed on the screen to place an empty cup, whose diameter