Maintaining Your Analyzer
10-16
Figure 10-20 Clean bath
10.3.10 Drain Bath
When at three or more of the WBC, RBC, PLT and HGB results are abnormal, you may do
the “
Drain bath
” procedure to find out the reason.
Follow the steps below to do so:
1. Do steps 1 to 4 of the “
Drain bath
” procedure to expose the bath;
2.
SELECT
“
Drain bath
” to drain the bath;
Figure 10-21 Draining the bath
3. When the draining is done, the screen will display the residual time of draining bath;