Using the Calibration Programs
9-11
SELECT
“
Calibration
→
Fresh Blood
” (Figure 9-10) to enter the “
Fresh Blood
” screen
(Figure 9-11).
Figure 9-11
“
Fresh Blood
”
screen
Complete the fresh blood calibration as instructed below:
Determination of the reference values
Reference values may be determined using three or more normal blood samples prepared for
calibration from the reference method or from a reliably calibrated hematology analyzer.
Editing calibration settings
Selecting
count
mode
At “
Fresh Blood
” screen, press [F1] to select the “
Whole Blood
” or “
Prediluted
” mode.
Selecting
sample
When the mode is selected, press [F3] to choose the fresh blood sample whose reference
values you want to set.
Editing the reference value
Press [F2] to enable the edit boxes in the “
Mean
” column, as Figure 9-12 shows.
Summary of Contents for BC-2800