Page 56
scanCONTROL Output Unit
scanCONTROL 29xx
7.4.4
OU-AnalogOut/4-Channel/0-10 V
- Micro-Epsilon order no. 0325135
- 4-Channel analog output module DC 0-10 V
- Indicators (Status-LED’s, for detailed description refer to manual of
the 4-Channel analog output module DC 0-10 V)
- Voltage range is terminal-related 0 to +10 V and can not be
extended by changing the appropriate setting in scanCONTROL
Configuration Tools.
- The output signal is galvanically isolated output from the system
level.
- The resolution is limited to 12 bits.
- scanCONTROL 2910/2960 supports four analog outputs.
Fig. 35 4-Channel-AnalogOut-Modul
Number of outputs
4
Power supply
via system voltage DC/DC
Signal voltage
0 V ... 10 V
Load
> 5 kΩ
Resolution
12 bit
Conversion time
typ. 10 ms
Setting time
typ. 100 μs
Measurement error at 25 °C
<± 0.1 % of full scale (FS)
Temperature coefficient
<± 0.01 % /K of full scale (FS)
Fig. 36 Technical data of analog output module
Summary of Contents for 4350006.71
Page 12: ...Page 12 Laser Safety thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 62: ...Page 62 Instruction Manual Software thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 64: ...Page 64 Pneumatic Plan thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 70: ...Page 70 Declaration of Incorporation thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 95: ...Operating Instructions scanCONTROL 29xx ...
Page 100: ......
Page 135: ...Page 41 Operation of the Sensor with a PC scanCONTROL 29xx Fig 24 Predefined measuring fields ...
Page 193: ......
Page 194: ......
Page 229: ......