Page 68
Maintenance, Service, Spare Parts Lists, Software Error Messages
thicknessCONTROL MTS 8201.LLT / 4350006.71
7.3
Software Error Messages
Error message
Actions
Sensor arm is not on top
-
Check free movement
-
Check emergency off
-
Wait for active movement
-
Check light barrier (alignment, free path of beam)
Sensor arm is not on bottom
-
Check free movement
-
Check emergency off
-
Wait for active movement
-
Check light barrier (alignment, free path of beam)
Traversing unit is not in zero position
-
Check free movement of the carriages
-
Wait for active movement of the unit
Calibration part not in position
-
Check free movement of pivoting arm
-
Check emergency off
-
Check light barrier (alignment, free path of beam)
-
Check compressed air
Very noisy sensor signals !!! - Please
check and clean the measuring sensors
-
Check ambient light
-
Check for contamination of the sensor glasses on
inside and outside
Wait for release of traversing
-
Wait for time x until the sensor head is in position
Misalignment of both axles - please check
their positions. Measurement accuracy may
no longer be achieved!!!
-
Check distances of carriages to limit switches in hom-
ing position
-
Check marks on the limit switches for misalignment
Test travel canceled - sensor gap is not
empty!!!
-
Check roller for contamination
-
Check measuring gap or roller for material
-
Stop button in software was pressed
Deviation of gap value from factory set-
tings is too large!
-
Check calibration part and sensors for contamination
-
Check position of laser line on calibration part (cen-
tered?)
-
If ambient temperature is cold: Let sensors warm up
Motor of calibration part does not rotate
- please check
-
Check supply voltage according to E plan
-
Check for free movement
-
Check for running noise
-
Check for mechanical fastening
Sensors at risk. Measurement stopped
-
Check light barrier (alignment, free path of beam)
-
Check log data for cause of error
-
Is signal via interface active?
-
Check for contamination of the sensor glasses on
inside and outside
Motor error Baumüller or EncoTRive
-
Check supply voltage
-
Check error codes on the controller
Summary of Contents for 4350006.71
Page 12: ...Page 12 Laser Safety thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 62: ...Page 62 Instruction Manual Software thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 64: ...Page 64 Pneumatic Plan thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 70: ...Page 70 Declaration of Incorporation thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 95: ...Operating Instructions scanCONTROL 29xx ...
Page 100: ......
Page 135: ...Page 41 Operation of the Sensor with a PC scanCONTROL 29xx Fig 24 Predefined measuring fields ...
Page 193: ......
Page 194: ......
Page 229: ......