Page 63
Operation
English
capaNCDT 6100
5.
Operation
5.1
Starting Up
Connect the display/output devices through the signal output socket, see Chap.
, before connecting
the device to the power supply and switching on the power supply.
i
Allow the measuring system to warm up for about one minute before the first measurement or calibrati-
on.
5.2
Basic Settings
The gain and zero point of the measuring channel are adjusted with the ’gain’ and ’zero’ trimmer potentiome-
ters, see
, (setting range approx. 20 turns per potentiometer).
The end settings at the left and right stops are recogniceable by a slight click.
Gain
Sensor/
Synchronization
Zero
Power/
Output
Fig. 7 Signal conditioning board with trimmer potentiometers