Page 28
Installation and Initial Operation
thicknessCONTROL MTS 8201.LLT / 4350006.71
4.4
Overview of Interfaces
1 .. 6
7 .. 8
Fig. 29 Interface connections required from the system operator
1
120V 60Hz / 10A
5
4x Analog output
2
Emergency stop line
6
Grounding
3
Remote support (VPN)
7
Grounding
4
Interface: Ethernet IP
8
air
Not supplied
4.5
Electrical Power Supply
The measuring system requires an AC mains voltage of 120 V, ±5 % / 60 Hz.
Connect the measuring system to the supply voltage.
The measuring system must be connected as follows:
Connect the cable for the 120V / 60Hz supply to -1X0:
Signal name
-1X0
contact
Customer
cable
L1 (120V ac)
1
N
2
PE
4
Connect the cable for the emergency stop line to -7X1:
Signal name
-7X1
contact
Customer
cable
emergency stop
contact : 1
1
emergency stop
contact : 1
2