6.4.3
Automatic Exposure Time Regulation
The automatic exposure time regulation facilitates the recording of the profile with optimum exposure time
(shutter time).
The
Shutter time
preset by the user is used as the starting value for the automatic exposure time regula-
tion. It should be selected so large that still valid measured values can be output at the darkest place of the
pro-file. One of the supplied demonstration programs can be used for determining this. The automatic expo-
sure time regulation can also be deactivated if required.
If there is no object in the measuring range, the “shutter time” stored in the shutter register of the sensor will
be used as exposure time. This value must also guarantee reliable recognition of the darkest measurement
object.
A previous test is recommended if the target is very dark or has very high contrast. The various demonstra-
tion programs are suitable for adjusting and testing the exposure time. In doing so, it is by all means sensible
to work with several different exposure times and to observe the effect in the diagrams.
The current exposure time can be calculated from the timestamps of the measured values. It is displayed as
Shutter time, see
Summary of Contents for 4350006.71
Page 12: ...Page 12 Laser Safety thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 62: ...Page 62 Instruction Manual Software thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 64: ...Page 64 Pneumatic Plan thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 70: ...Page 70 Declaration of Incorporation thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 95: ...Operating Instructions scanCONTROL 29xx ...
Page 100: ......
Page 135: ...Page 41 Operation of the Sensor with a PC scanCONTROL 29xx Fig 24 Predefined measuring fields ...
Page 193: ......
Page 194: ......
Page 229: ......